品質 Karl Storz 27005BA Cystoscope for repair 工場

Karl Storz 27005BA Cystoscope for repair

Karl Storz 27005BA Cystoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
品質 LINVATEC HD 4300 Arthroscope for repair 工場

LINVATEC HD 4300 Arthroscope for repair

LINVATEC HD 4300 Arthroscope for repair ► Brief Description Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Solution: Replace objective lens, inner optical parts, clean and adjust optical system. ►Problem & Solutoin Brand LINVATEC Model HD 4300 Fault Image with ...
品質 Stryker 30º 0502-503-030 laproscope for repair 工場

Stryker 30º 0502-503-030 laproscope for repair

Stryker 30º 0502-503-030 laproscope for repair ► Brief Description Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. ►Problem & Solutoin Brand ...
品質 オリンパス A70940A 修理用オートクレーブ対応関節鏡 工場

オリンパス A70940A 修理用オートクレーブ対応関節鏡

Olympus A70940A Autoclavable Arthroscope for repair ► Brief Description Fault: 1. Image blur, black dots and dark. 2. Fiber not light enough. 3. Rod lens broken. 4. Objective lens broken and dirty. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...
品質 修理用オリンパスGIF-Q150胃内視鏡 工場

修理用オリンパスGIF-Q150胃内視鏡

Olympus GIF-Q150 Gastroscope for repair ► Brief Description Fault: 1. With image; 2. Light bundlebreak badly; 3. Insert tubeaging; 4. Biopsy tubeabrasion, leakage badly; 5. Angle tied, bendingsection, wires and helix tube abrasion; 6. Rubberwithout glue. Solution: Replace light bundle, insert tube, ...
品質 GIF-XQ240 オリンプスのガストロスコップ 修理のため 工場

GIF-XQ240 オリンプスのガストロスコップ 修理のため

GIF-XQ240 Gastroscope by Olympus for repair ► Brief Description Fault: Leakage. Solution: Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ►Problem & Solutoin Brand Olympus Model GIF-XQ240 Fault Leakage. Solution Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ► Common ...
品質 オリンパス CYF V2 修復用のシストスコップ 工場

オリンパス CYF V2 修復用のシストスコップ

Olympus CYF V2 Cystoscope For Repair ► Brief Description Fault: Leakage. Solution: Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber. ►Problem & Solutoin Brand Olympus Model CYF V2 Fault Leakage Solution Replace the insert tube, light bundle, biopsy ...
品質 オリンパス GIF-H170 胃内視鏡 修理用 工場

オリンパス GIF-H170 胃内視鏡 修理用

Olympus GIF-H170 gastroscope for repair ► Brief Description Fault: 1. The top left and middle right side with blur image, objective Lens broken. 2. 20% light bundle broken, the small lens turned yellow. 3. Insert tube yellow and wrinkles. 4. The biopsy tube abrasion. 5. The wire and helix tube ...
品質 R ウォルフ 7326071 修復のための尿道鏡 工場

R ウォルフ 7326071 修復のための尿道鏡

R Wolf 7326071 ureteroscope for repair ► Brief Description Fault: The image has spots,image bundle aging,light bundle broken;Insert tube aging;Biopsy channel worn and leakage;Bending section,wire and helix tube were worn. Solution: Replace the light bundle,image bundle,rubber,biopsy channel,helix ...
品質 インシスター IS3-F喉鏡 修理のために 工場

インシスター IS3-F喉鏡 修理のために

Insighters IS3-F laryngoscope for repair ► Brief Description Fault: 1. Image blank screen; 2. Insert tube abrasion 3. Bending section transformation; 4. Distal head transformation; Solution: Replace CCD, insert tube, bending section, rubber ►Problem & Solutoin Brand Insighters Model IS3-F Fault 1. ...
品質 KARL STORZ 11278A1 尿管鏡(修理用) 工場

KARL STORZ 11278A1 尿管鏡(修理用)

KARL STORZ 11278A1 ureteroscope for repair ► Brief Description Fault: 1) Image spotted with partially black point, image bundle aged 2) Light bundle partially wire broken 3) Insert tube aged 4) Biopsy tube abrasion and leakage 5) Wires and coil pipe abrasion Solution: Replace HD image bundle, light ...
品質 オリンパス ENF-GP 繊維鼻喉鏡 修理用 工場

オリンパス ENF-GP 繊維鼻喉鏡 修理用

Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair ► Brief Description Fault: 1. Image with black point and image bundle broken 2. Light bundle aged 3. Insert tube aging and wrinkle 4. Bending section deformation 5. Angle with not enough 6. Rubber abrasion Solution: Replace image bundle; ...
品質 オリンパスBF-XP160F 器具修理 工場

オリンパスBF-XP160F 器具修理

Olympus BF-XP160F fiber scope for repair ► Brief Description Fault: 1. Image degumming, serious color, CCD aging, a few black and gray spots; 2. 20% of the light bundle is broken; 3. Insert tube aging slightly; 4. Biopsy tube wear, not smooth; 5. Bending section, steel wire, screw tube wear; ...
品質 オリンパス CF-2T160L 大腸内視鏡 修理用 工場

オリンパス CF-2T160L 大腸内視鏡 修理用

Olympus CF-2T160L Colonoscope for repair ► Brief Description Fault: 1. With image. 2. 10% light bundle broken. 3. Insert tube is ok. 4. Biopsy tube abrasion. 5. Angle tied, bending section transformation. 6. Wires, helix tube abrasion, no up angle. Solution: Replace image bundle; insert tube; biopsy ...
品質 オリンパス CF-Q160AL 大腸内視鏡 修理用 工場

オリンパス CF-Q160AL 大腸内視鏡 修理用

Olympus CF-Q160AL scope for repair ► Brief Description Fault: 1.No image, CCD broken; 2.image broken completely; 3.insert tube aging; 4.biopsy tube abrasion; 5.bending section abrasion; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace CCD, light bundle, insert ...
品質 オリンパスCHF-P20Q内視鏡 修理用 工場

オリンパスCHF-P20Q内視鏡 修理用

Olympus CHF-P20Q endoscope for repair ► Brief Description Fault: 1.Insert tube cracked; 2.Rubber leakage; 3.Image with black point and cracked; 4.Angulation tight; 5.Biopsy tube aging Solution: Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber ►Problem & ...
品質 オリンパスGIF-1TQ160 ガストロスコップ 修理 工場

オリンパスGIF-1TQ160 ガストロスコップ 修理

Olympus GIF-1TQ160 gastroscope for repair ► Brief Description Fault: 1.With image, but CCD with broken. 2.Light bundle broken completely. 3.Insert tube abrasion. 4. No biopsy tube. 5. No air&water tube, no water supply tube. 6. No bending section assembly. 7. No digital tip. Solution: Replace CCD, ...
品質 オリンパスGIF-H180J ガストロスコップ 修理 工場

オリンパスGIF-H180J ガストロスコップ 修理

Olympus GIF-H180J gastroscope for repair ► Brief Description Fault: 1.image with shadow; 2.light bundle broken; 3.insert tube with bite mark; 4.biopsy tube abrasion; 5.bending section serious deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace CCD, ...
品質 オリンパスGIF-Q180 ガストロスコップ 修理 工場

オリンパスGIF-Q180 ガストロスコップ 修理

Olympus GIF-Q180 gastroscope for repair ► Brief Description Fault: 1.image definition not enough, CCD aging. 2.30% light bundle broken. 3. Insert tube aging and wrinkle. 4.Biopsy tube abrasion. 5. Bending section bited, transformation. 6. Angle not enough, wires and helix tube abrasion. 7. Light ...
品質 オリンパス CF-240AI コロノスコップの修理 工場

オリンパス CF-240AI コロノスコップの修理

Olympus CF-240AI Colonoscope for repair ► Brief Description Fault: Image blur,insert tube aging,light guide squashed,light guide broken 50%,nozzle blocked,angle tight,forceps channel aging. Solution: Replace insertion tube,CCD chips,light guide,forceps channel,water channel,angulation wire,bending ...
品質 Olympus LTF-160 胸腔鏡修理用 工場

Olympus LTF-160 胸腔鏡修理用

Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the ...
品質 オリンパス TJF-160VR 内視鏡 修理 工場

オリンパス TJF-160VR 内視鏡 修理

Olympus TJF-160VR endoscope for repair ► Brief Description Fault: 1.No image, CCD broken. 2. Light bundle broken completely. 3. Insert tube aging and wrinkle. 4. Biopsy tube broken. 5. Wires and helix tube broken. 6. clamp wires broken. 7. Operation part broken, no angle. 8. Grip broken. 9. The ...
品質 修理用オリンパスGIF-Q150胃内視鏡 工場

修理用オリンパスGIF-Q150胃内視鏡

Olympus GIF-Q150 gastroscope for repair ► Brief Description Fault: 1.with image; 2.one eye of light bundle broken; 3.insert tube little aging; 4.biopsy tube abrasion; 5.bending section deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle. Solution: Replace light bundle ...
品質 STORZ 13820pks 胃内視鏡 修理用 工場

STORZ 13820pks 胃内視鏡 修理用

STORZ 13820pks Gastroscope for Repair ► Brief Description Fault: 1.Rubber leakage; 2.Wire abrasion; 3. no up angle, other angle lack of. Solution: 1. Replace the wires, coil pipe, rubber; 2.Adjust angle. ►Problem & Solutoin Brand STORZ Model 13820pks Fault 1.Rubber leakage; 2.Wire abrasion; 3. no up ...